Curve Tracer
371A
This product is discontinued.
Characteristics
Collector Supply
Modes - High Current: 250 µs
pulses with max. peak of 30 volts. High Voltage: full
rectified sine with max. peak of 3000 volts (positive and negative
polarities for both modes).
|
Range |
Peak
Current |
Peak
Power |
High Current |
30 V |
400 A |
3 kW |
(pulsed) |
30 V |
40 A |
300 W |
High Voltage |
3 kV |
40 mA
±20% |
30 W |
|
3 kV |
4 mA
±20% |
3 W |
|
300 V |
4 mA
±20% |
300
mW |
|
300 V |
0.4 mA ±20% |
30
mW |
Measurement Characteristics
Collector Current
Maximum
Power Setting |
Vertical Range (in 1-2-5 increments) |
3 kW |
1 A/div to 50 A/div |
300 W |
500 mA/div to 5 A/div |
30 W |
100 µA/div to 5 mA/div |
3 W, 300
mW |
10 µA/div
to 500 µA/div |
30 mW |
1 µA/div to
50 µA/div |
Test Fixture - Designed to allow easy
connection to a variety of devices. It has a safety enclosure to assure
operator protection. Special patch cords are provided for connecting large
devices.
Step Generator
Modes - Current (stair step): DC,
500 µs ±10% of pulse. Voltage (stair step): DC.
Step Range - Current: 1 mA/step
to 2 A/step pulsed (high current mode). 1 µA/step to
2 mA/step DC (high voltage mode). Voltage: 200 mV/step
to 5 V/step. Offset: ±10X step amplitude (5X in
5 V/step and 2 V/step settings).
Safety - UL1244, CSA231,
HD401 S1 |